User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Cyber-Systems FG: Function Generator Module DATE: 06-Mar-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 80% NUMBER OF TESTS: 12 NUMBER OF LINES: 192 CONFIGURATION: Datron 1281 CONFIGURATION: HP 5335A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP Set the UUT front panel controls as follows: 1.004 DISP 1.004 DISP FAST/SLOW switch ................... FAST 1.004 DISP NORMAL/INVERT switch ............... NORMAL 1.005 DISP Set the FUNCTION CONTROL MODULE front panel controls 1.005 DISP as follows: 1.005 DISP ACTIVE/DISABLE switch .............. ACTIVE 1.005 DISP TOTAL/PRESET switch ................ PRESET 1.005 DISP A/B switch ......................... A 1.005 DISP INTERNAL/EXTERNAL FUNCTION switch .. INTERNAL 1.005 DISP INTERNAL SOFT START/STOP ........... OFF (down) 1.006 DISP Set the FUNCTION CONTROL MODULE front panel controls 1.006 DISP as follows: 1.006 DISP FUNCTION WAVEFORM .................. SINE 1.006 DISP MULTIPLIER ......................... X10 1.006 DISP FREQUENCY .......................... 10 1.006 DISP MASTER SPAN ........................ MAX 1.007 HEAD {} 1.008 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.009 HEAD {º DAC CALIBRATION º} 1.010 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.011 HEAD {}ÍÍ DAC CALIBRATION ÍÍ 1.012 JMP 2.004 1.013 EVAL 2.001 IEEE [@1281] DCV AUTO 2.002 DISP Adjust R13 for a 1281 reading of -10.000V. 2.003 JMP 2.009 2.004 DISP Apply power to the main frame. Use the FCM CURSOR and 2.004 DISP INCREM switches to set the display to 00000000 and 2.004 DISP set the TOTAL/PRESET switch to TOLTAL position. 2.005 DISP Connect the UUT OUTPUT BNC connector to the 1281 2.005 DISP VOLTS INPUT. 2.006 DISP On main frame, set the test switches as follows: 2.006 DISP 2.006 DISP 90ø ....................... OPEN 2.006 DISP 180ø ...................... OPEN 2.006 DISP 270ø ...................... SHORT 2.007 DISP Press and hold UUT RUN/PAUSE for at least 5 seconds. 2.008 DISP Release to pause state (indicator flashing). 2.009 IEEE [@1281] DCV AUTO 2.010 IEEE [@1281] RDG?[I] 2.011 MATH MEM1 = -10 2.012 MEME 2.013 MEMC V 0.004U 3.001 DISP On main frame, set the test switches as follows: 3.001 DISP 3.001 DISP 90ø ....................... SHORT 3.001 DISP 180ø ...................... OPEN 3.001 DISP 270ø ...................... OPEN 3.002 JMP 4.004 3.003 EVAL 4.001 IEEE [@1281] DCV AUTO 4.002 DISP Adjust R11 for a 1281 reading of 10.000V. 4.003 JMP 4.006 4.004 DISP Press and hold UUT RUN/PAUSE for at least 5 seconds. 4.005 DISP Release to pause state (indicator flashing). 4.006 IEEE [@1281] DCV AUTO 4.007 IEEE [@1281] RDG?[I] 4.008 MATH MEM1 = 10 4.009 MEME 4.010 MEMC V 0.004U 5.001 HEAD {} 5.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 5.003 HEAD {º INVERTED OUTPUT CALIBRATION º} 5.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 5.005 HEAD {}ÍÍ INVERTED OUTPUT CALIBRATION ÍÍ 5.006 DISP On main frame, set the test switches as follows: 5.006 DISP 5.006 DISP 90ø ....................... OPEN 5.006 DISP 180ø ...................... SHORT 5.006 DISP 270ø ...................... OPEN 5.007 DISP Press and hold UUT RUN/PAUSE for at least 5 seconds. 5.008 DISP Release to pause state (indicator flashing). 5.009 IEEE [@1281] DCV AUTO 5.010 IEEE [D3000] 5.011 IEEE [@1281] RDG?[I] 5.012 MEM* 1000 5.013 MATH M[1] = MEM 5.014 DISP Set the UUT front panel controls as follows: 5.014 DISP 5.014 DISP NORMAL/INVERT switch ................ INVERT 5.015 MATH MEM1 = M[1] 5.016 DISP Adjust R15 as necessary to obtain a 1281 reading 5.016 DISP equal to in magnitude but opposite in sign as [MEM1]V. 5.017 IEEE [@1281] DCV AUTO 5.018 IEEE [@1281] RDG?[I] 5.019 MEM* 1000 5.020 MATH M[2] = MEM 5.021 MATH MEM1 = M[1] 5.022 RSLT =[MEM1]mV = Normal Signal 5.023 RSLT = 5.024 MATH MEM1 = M[2] 5.025 RSLT =[MEM1]mV = Inverted Signal 5.026 HEAD {}ÍÍ INVERTED OUTPUT CALIBRATION ÍÍ 5.027 DISP On main frame, set the test switches as follows: 5.027 DISP 5.027 DISP 90ø ....................... OPEN 5.027 DISP 180ø ...................... OPEN 5.027 DISP 270ø ...................... SHORT 5.028 JMP 6.004 5.029 EVAL 6.001 IEEE [@1281] DCV AUTO 6.002 DISP Adjust R23 for a 1281 reading of 10.000V. 6.003 JMP 6.006 6.004 DISP Press and hold UUT RUN/PAUSE for at least 5 seconds. 6.005 DISP Release to pause state (indicator flashing). 6.006 IEEE [@1281] DCV AUTO 6.007 IEEE [@1281] RDG?[I] 6.008 MATH MEM1 = 10 6.009 MEME 6.010 MEMC V 0.004U 7.001 DISP Set the main frame power to the off position. 7.002 DISP Disconnect the test set-up. 7.003 HEAD {} 7.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 7.005 HEAD {º PHASE ACCURACY TEST º} 7.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 7.007 HEAD {}ÍÍ PHASE ACCURACY TEST ÍÍ 7.008 ASK- F 7.009 JMP 8.001 7.010 EVAL 8.001 DISP Set the UUT front panel controls as follows: 8.001 DISP 8.001 DISP NORMAL/INVERT switch ................ NORMAL 8.002 DISP On main frame, set the test switches as follows: 8.002 DISP 8.002 DISP 90ø ....................... OPEN 8.002 DISP 180ø ...................... OPEN 8.002 DISP 270ø ...................... OPEN 8.003 DISP Install second Function Generator Module in main frame 8.003 DISP slot right of UUT. Set unit front panel controls the 8.003 DISP same as the UUT. 8.004 DISP Set the FUNCTION CONTROL MODULE front panel controls 8.004 DISP as follows: 8.004 DISP 8.004 DISP FUNCTION WAVEFORM ......... SQUARE 8.004 DISP MASTER SPAN ............... slightly cw from MIN 8.005 DISP Connect the UUT OUTPUT BNC connector to the 8.005 DISP 5335A INPUT A. 8.006 DISP Connect the OUTPUT BNC on the Function Generator Module 8.006 DISP in the second main frame slot to the 5335A INPUT B. 8.007 DISP Set the main frame power to the on position. 8.008 IEEE [@5335] [TERM LF] 8.009 IEEE [@5335] IN[D2000]; 8.010 IEEE [@5335] FN16,AF1,BF1; 8.011 DISP Press and hold UUT RUN/PAUSE for at least 5 seconds. 8.011 DISP Release to pause state (indicator lit). 8.012 IEEE [@5335] [D1000][I] 8.013 MATH MEM1 = 0 8.014 MEME 8.015 MEMC ø 1.0U 9.001 DISP Adjust the UUT UP/DOWN PHASE LEAD control for 9.001 DISP a UUT indication of 90ø. 9.002 IEEE [@5335] [D1000][I] 9.003 MATH MEM1 = 90 9.004 MEME 9.005 MEMC ø 1.0U 10.001 DISP Adjust the UUT UP/DOWN PHASE LEAD control for 10.001 DISP a UUT indication of 180ø. 10.002 IEEE [@5335] [D1000][I] 10.003 MATH MEM1 = 180 10.004 MEME 10.005 MEMC ø 1.0U 11.001 DISP Adjust the UUT UP/DOWN PHASE LEAD control for 11.001 DISP a UUT indication of 270ø. 11.002 IEEE [@5335] [D1000][I] 11.003 MATH MEM1 = 270 11.004 MEME 11.005 MEMC ø 1.0U 12.001 DISP Set the main frame power to the off position. 12.002 DISP Disconnect the test set-up. 12.003 HEAD {} 12.004 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 12.004 DISP {º THIS COMPLETES THE VERIFICATION º} 12.004 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 12.005 END